Abstract

The use of a liquid sampling-atmospheric pressure glow discharge (LS-APGD) microplasma source as an alternative to conventional inductively coupled plasma (ICP) detection of laser ablation (LA) produced particles using a Nd:YAG laser at 1064nm is demonstrated. This configuration utilizes a 180° geometry, which is different from the 40° geometry that was used to ionize ablated particles followed by mass spectrometric detection. The use of a hollow counter electrode (nickel, 0.3cm o.d., 0.1cm i.d.) was implemented to introduce ablated particles directly into the APGD plasma with helium as a carrier gas. The LS-APGD source was optimized using ablated copper as the test sample (helium carrier gas flow rate (0.30Lmin−1 He), discharge current (60mA), laser power (44mJ), and solution electrode sheath gas (0.2Lmin−1 He) and solution flow rates (10μLmin−1 5% HNO3)). Standard brass samples having known Zn/Cu percentages were ablated and analyzed using the LS-APGD source. As a comparison, the established technique of laser-induced breakdown spectroscopy (LIBS) was used to analyze the same set of brass standards under similar ablation conditions to the LS-AGPD measurements, yielding comparable results. The Zn/Cu ratio results for the LS-APGD and LIBS measurements showed good similarity to previous measurements using ICP-MS detection. The performance of the LS-APGD–OES microplasma, comparable to well established methods, with lower capital and operational overhead expenses, suggests a great deal of promise as an analytical excitation source.