Abstract

An isocratic liquid chromatographic method for the assay and purity control of josamycin using wide-pore poly(styrene-divinylbenzene) as the stationary phase is presented. Josamycin has been well separated from all the related leucomycins, present as impurities in commercial samples. Wide-pore PLRP-S 8 μm 1000 A has shown the best selectivity. The separation has been carried out at 65°C with a mobile phase of 2-methyl-2-propanol/0.2 mol/l potassium phosphate buffer pH 10.5/water (24.5: 5: 70.5, V/V) and a flow rate of 1.0 ml/min. UV detection has been performed at 232 nm. The total analysis time has been about 40 min. The method has been used to analyse commercial bulk samples and tablets containing josamycin base.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.