Abstract

A liquid chromatographic method for analysis of tylosin A and related substances was developed using wide-pore poly(styrene-divinylbenzene) as the stationary phase. As a mobile phase, a mixture of tetrahydrofuran-0.2 M potassium phosphate buffer pH 9.0-water (20:5:75) was used at a flow-rate of 1.0 ml/min. The column was kept at 60°C. Detection was performed by UV at 280 nm. Tylosin A is separated from all its known related substances that are potentially present as impurities in commercial samples of tylosin base and tylosin tartrate. The method shows good linearity and repeatability and is stability indicating. Results for a number of commercial samples analysed using the newly developed method are presented.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.