Abstract

AbstractTrace enrichment for the GC analysis of a series of chlorinated pesticides and polychlorinated biphenyls (PCBs) in aqueous samples has been achieved through a simple on‐line technique involving sorption on an LC micro‐precolumn followed by direct elution into a gas chromatograph with hexane. A 5‐m retention gap coupled to the capillary GC column served as the recipient of a relatively large sample volume (ca. 100 μl) introduced into the GC. Partially concurrent solvent evaporation during sample introduction allowed a large sample capacity. Recoveries of more than 95% were observed for the majority of the compounds studied. Using 1.0 ml aqueous samples, detection limits of less than 1 ppt were found. The applicability of the developed method was demonstrated for a river water sample.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.