Abstract
Di‐2‐pyridyl ketone thiosemicarbazone (DPKT) was examined as a precolumn chelating reagent for high performance liquid chromatographic separation of Co(II), Cu(II), Ni(II), and Fe(II) ions as metal chelates on a 10 µm Bondapak C18 (300 mm×3.9 mm) RP column. A gradient program with a mobile phase of acetonitrile and 0.01 M pH:6.0 acetate buffer is used. Acetonitrile composition is increased from 25 to 40% at 18 min and held at 40% for 18 min. A UV detector was used at 400 nm. The flow rate was 1.5 mL · min−1. Limit of detection was determined on a 3σ blank basis. Interference effects from other metal ions were also investigated. The method was applied for the determination of Co(II), Cu(II), Ni(II), and Fe(II) ions in drinking and tap water samples.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Liquid Chromatography & Related Technologies
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.