Abstract

Intraocular pressure (IOP) fluctuation is an independent risk factor for glaucoma and a key to clarify its pathogenesis. We have developed a contact lens integrated with a conformal LC resonance circuit made of liquid alloy (Galinstan) for non-invasive continuous IOP monitoring. The contact lens is fabricated using polydimethylsiloxane (PDMS), in which the LC resonance circuit can sense corneal strain proportionally caused by IOP variation. The Galinstan configuration is formed to the designed pattern by UV-laser treating the carbon-doped Ecoflex (c-Ecoflex) stamps and then selectively transferring the Galinstan onto the stamp to form the LC resonance circuit pattern. The tests conducted on silicone model eyes demonstrate high sensitivity in the frequency responses to variations in IOP, reaching 649 ppm/mmHg. IOP measurement on human eyes during head-down tilt bedrest confirms its applicability. With process optimization, this contact lens sensor has great potential for mass production and wide application in monitoring IOP.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.