Abstract
ABSTRACT Test made on four ITT FW-130 photomultipliers in the pulse-counting mode show that they are linear with respect to an RCA 1P21 to better than 0.002 magnitude per magnitude over a 7.5-magnitude range in brightness. Key words: photometry-photomultipliers