Abstract

Some of the more common modes of spectrograph slit irradiation have been studied experimentally. Particular attention has been paid to line shape and its effect on resolution. The noncoherent mode is seldom if ever employed because of practical limitations. The lens and broad source modes offer a means of realizing high flux transmission, but the consequent line resolution may be poor. The resolution can be improved by reducing either the relative aperture of the condenser lens or the angle subtended by the broad source at the slit. The limit in either case is the coherent mode with which very sharp, well-resolved lines can be obtained. The closest approach to flat top line shape is realized through the use of a wide slit when the departure from noncoherent irradiation is not very great.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.