Abstract

The intraoperative shift of neuroanatomical landmarks that serve as reference points is an unsolved problem in current neuronavigation. Monitoring the position of these landmarks requires repeated intraoperative imaging. We analyzed the effectiveness of a 3-D ultrasound system for imaging through a bore-hole trepanation. A tissue-mimicking ultrasound phantom and plastic pads with bore-holes were utilized for in vitro tests of the system. Reducing the diameter of the simulated bore-hole decreased the image quality and reduced the field of view. The multiple plane mode of the 3-D ultrasound system allows reconstruction of images in arbitrary imaging planes on the basis of intraoperatively acquired 3-D datasets. Selecting planes that are coplanar with preoperative MRI scans, we were able to identify neuroanatomical landmarks in the reconstructed ultrasound images. Repeated 3-D ultrasound during a procedure might, therefore, allow recognition of a shift of these landmarks.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.