Abstract
Calibration-free laser-induced breakdown spectroscopy (CF-LIBS) was applied to tungsten (W)- tantalum (Ta) alloy and W-rhenium (Re) alloy. In the W-Ta alloy measurement, an apparent Ta I line at 648.5 nm was found to be isolated from W I lines and used for the concentration determination. Accurate concentration measurements were achieved for a W - 2 at.% Ta specimen at the ambient gas pressure between 0.1 and 10 Torr of He, Ne and Ar gas. Within these ambient gas species, the highest line intensities were obtained by Ar gas. In the W-Re alloy measurement, the detection limit of trace Re in W-Re was investigated. CF-LIBS results agree well with the nominal concentration of Re down to W-0.5 at.% Re specimen which is better than the measurement accuracy of EDX (energy-dispersive X-ray micro-analysis). Further improvements are required in order to measure 0.2 at.% Re in W expected at ITER.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.