Abstract

The extraction of the rabbit lens is described using a 25 G irrigating needle and a 22 G aspirating needle; at the latter's bevelled tip lens fragmentation occurs due to the longitudinal ultrasonic vibrations generated there--an 'acoustic horn' causes the tip to vibrate with large amplitudes. The use of small needles allows considerable manoeuvrability in the anterior chamber and usually eliminates the need for corneal suturing. Push-pull coupled syringes equate the volume of irrigation with that of aspiration. This procedure makes possible lens extraction through an aperture in the anterior capsule of the rabbit's lens and a similar machine is being constructed for trial on human cataract.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.