Abstract

Analytical method development and validation for determination of favipiravir (FVPR) in pure and tablet dosage forms by liquid chromatography with tandem mass spectrometry/mass spectrometry (LC-MS/MS) technique. A simple LC-MS/MS method was developed for determination of a new antiviral drug, FVPR in pharmaceutical formulations. The stationary phase employed was a Shim pack GISS, C18 (100 mm × 2.1 mm, 1.9 μm) column and mobile phase used in pump A was 10.0 mM ammonium acetate and in pump B methanol was used. The gradient program was used with fixed mobile phase flow rate at 0.4 mL min-1. Total run time was 5.0 min. The proposed method was validated according to International Conference on Harmonization (ICH) guidelines. The established method found better outcomes. The linearity graph was found in the range of 50-200 μg/mL and the correlation coefficient value (R2) obtained was found to be 1.0. The limit of detection (LOD) and limit of quantification (LOQ) were 4.044 μg/mL and 12.253 μg/mL, respectively. Tremendous recovery outcomes were observed and found to be 101%, 99.0%, and 99.5% for FVPR at 150% upper, 100% middle, and 50% lower concentrations, respectively. All outcomes obtained comply with ICH guidelines. The developed method was simple, unique, accurate, robust, precise, and reproducible for determination of FVPR in tablet formulation. The method is novel and could be adopted in formulation industry.