Abstract

A precise calibration and monitoring system has been developed for the HARP experiment scintillator-based time of flight system. An Nd-YAG laser with passive Q-switch and active/passive mode-locking and a custom-made laser light injection system based on a bundle of IR monomode optical fibers were used. The laser pulse timing was provided by a novel ultrafast InGaAs MSM photodiode, with 30 ps risetime. Experience over a several month data taking period in 2001 and 2002 shows that drifts in timing down to about 70 ps can be traced.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.