Abstract

Large-volume sample stacking capillary electrophoresis (LVSS-CE) and conventional capillary electrophoresis (CE) are compared for the separation of drugs of significance to forensic and clinical analyses. LVSS-CE for cations requires the use of an electroosmotic flow (EOF) modifier in conjunction with polarity switching to effect on-column concentration of an analyte and its subsequent migration in the capillary. The run buffer consists of 0.05 mol dm −3 disodium tetraborate adjusted to pH 2.2 with orthophosphoric acid, and the EOF modifier is 0.002 mol dm −3 cetyltrimethylammonium bromide. CE investigations used an identical run buffer minus the EOF modifier. LVSS-CE and CE investigations used injection times of 30 s and 3 s, respectively. Both modes of capillary electrophoresis are compared in terms of their limits of detection, efficiency, resolution and reproducibility. LVSS-CE is also applied to the analysis of a spiked urine sample.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.