Abstract

BackgroundHighly cross-linked polyethylene (XLPE) is reported to have low rates of linear and volumetric wear at 10-14 years. In a prior study, larger (36 and 40 mm) femoral heads were associated with more volumetric wear, but there were only 12 hips with these heads. MethodsWe evaluated 107 hips (93 patients, with a mean age of 76 years) with one design of uncemented acetabular component, a 36 (90 hips) or 40 mm (17 hips) metal femoral head, and one electron beam 100 kGy irradiated and remelted XLPE at a mean follow-up of 8 years (range 5-13 years). Selection of these femoral heads was based on several factors, including the perceived risk of dislocation, the outer diameter size of the acetabular component, and liner availability. Measurements of linear and volumetric wear were performed in one experienced laboratory by the Martell method and analyzed using the first-to-last method. Standard radiographs, with additional Judet views, were used to detect periprosthetic osteolysis. Clinical records were used to determine the complications of dislocation, liner fracture, and painful trunnion corrosion. ResultsFor the entire cohort, the median linear wear rate was 0.041 mm/y (95% confidence interval, 0.031-0.057) and the median volumetric wear rate was 34.6 mm3/y (95% confidence interval, 31.4-53.5). With the numbers available, there was no difference in linear or volumetric wear between the 36 and 40 mm head sizes. Small, asymptomatic osteolytic lesions were noted in 3 hips (2%). There were 3 patients (3%) with dislocation (2 early and 1 late), but these have not had a revision. There were no revisions for loosening, no liner fracture, and no patient with symptomatic trunnion corrosion. ConclusionThis acetabular component and XLPE with large metal heads had low rates of linear and volumetric wear. Large metal femoral heads did not lead to liner fracture, loosening, or symptomatic trunnion corrosion in this patient population. However, we recommend longer clinical follow-up studies and caution in the routine use of larger metal femoral heads in other, younger patient populations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call