Abstract

The presented paper summarizes our earlier studies on application of beveled-edge Large Area Avalanche Photodiodes (LAAPDs) in γ-rays scintillation detection. LAAPDs, due to their high quantum efficiency and low excess noise factor allow for better statistical accuracy of the signal as compared to photomultipliers. The device dark noise contribution significantly affects energy resolution only for γ-rays with energy below 50 keV. Notably better or comparable energy resolutions to those observed with a XP2020Q photomultiplier were obtained with the LAAPDs for a number of different scintillators. Particularly, the recorded energy resolutions of 4.3±0.2% and 4.8±0.14% measured with YAP and CsI(Tl) crystals, respectively, for the 662 keV γ-peak from a 137Cs source belong to the best observed ever with these scintillation detectors. Results of the study of timing with fast scintillators coupled to the LAAPD showed subnanosecond time resolution of 570±30 ps for 60Co γ-rays detected in LSO crystal. The response of LAAPD to X-rays and factors limiting energy resolution have been discussed too.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.