Abstract

The Pro-Tek passive colorimetric air monitoring badge for personal or area sampling of hydrogen sulfide is described. The badge has been validated over the range of 1.8 to 164 ppm-hours (0.23-21 ppm on an 8-hour TWA basis). It has an overall accuracy throughout this range of +/- 15.9% and meets the NIOSH accuracy criteria for an analytical and sampling method. The colorimetric analytical method used is based on the Texas Air Control Board's Molybdenum Blue method. Color-activated exposed badge solutions are read out on a standard laboratory spectrophotometer using 1 centimeter (10 mm) cells. Variations in exposure temperature (between 10 degrees and 40 degrees C), relative humidity, and face velocity (between 2 and 250 ft/min) do not affect badge performance. Unexposed badges are stable for more than 12 months refrigerated and for two months at room temperature.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.