Abstract

C1 inhibitor deficiency may be hereditary or acquired. It is characterized by absent or poorly functioning C1 inhibitor. The disorder is rare, with prevalence estimated at 1/50,000. The very low incidence of the condition makes the sensitivity and specificity of assays used particularly important. Two different methods are commercially available to measure C1 inhibitor function. There are few data comparing these assays. Two assays of C1 inhibitor function (C1 inhibitor-C1s complex formation or inhibition of C1 esterase cleavage of artificial substrate [colorimetric]) were compared in 71 patients (28 hereditary angioedema, 2 acquired angioedema and 41 controls). Qualitatively, the two assays showed good correspondence (92%). Six of 71 results were discordant. Correlation in quantitative terms was moderate (R = 0.81). Both assays show high sensitivity for hereditary/acquired angioedema. The colorimetric assay is more prone to false-positive results. However, clinical interpretation is not adversely affected.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.