Abstract

We demonstrate for the first time that the detailed blood flow distribution within intracranial dura mater and cortex can be visualized by an ultrahigh sensitive optical microangiography (UHS-OMAG). The study uses an UHS-OMAG system operating at 1310 nm with an imaging speed at 150 frames per second that requires approximately 10 s to complete one 3-D scan of approximately 2.5 x 2.5 mm(2). The system is sensitive to blood flow with a velocity ranging from approximately 4 microms to approximately 23 mms. We show superior performance of UHS-OMAG in providing functional images of capillary level microcirculation within meninges in mice with the cranium left intact, the results of which correlate well with the standard dural histopathology.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.