Abstract

Using 3-mercaptopropyltrimethoxysilane (MPS)-coated (PbMg 1/3 Nb 2/3 O3)0.63-(PbTiO3)0.37 (PMN-PT)/tin and lead zirconate titanate/glass piezoelectric microcantilever sensors (PEMSs) with single-chain variable fragment (scFv) immobilized on the MPS surface, we have demonstrated real-time, label-free detection of human epidermal growth factor receptor 2 (Her2) in a background of 1 mg/ml bovine serum albumin. Coupled with a scFv with a KD of 3.4 x 10(-8)M, the MPS-insulated PMN-PT/tin PEMS 560 microm long and 720 microm wide exhibited a Her2 concentration sensitivity of 5 ng/ml in a background of 1 mg/ml BSA.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.