Abstract

We describe a new kinetic method for potentiometric determination of creatinine in serum, based on the creatinine--picrate reaction in alkaline medium (Jaffé reaction). The reaction is monitored with a picrate-selective electrode, and the increase in electrode potential during 270 s is measured and related directly to the creatinine concentration. Small cation-exchange columns are used to separate creatinine from interfering substances. Analytical recovery of creatinine added to serum was 100.7%. Results for a series of samples compared well with results obtained with a spectrophotometric method (r = 0.994).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.