Abstract

This paper presents a real-time functional decomposition adaptive algorithm for the optimal sampling of the interferometric signal in Swept-Source Optical Coherence Tomography imaging systems, which completely eliminates the input signal dependent nonlinearities that are problematic in current state-of-the-art OCT realizations that use interpolation and resampling. The proposed adaptive calibration algorithm uses the Kalman approach to estimate the wavenumber index parameter k from the Mach-Zender Interferometer signal which is then applied to an adaptive level crossing sampler to generate a sampling clock that k-linearizes the data on real-time during the sampling process. Such a system implements an artifact-free realization of the technology removing the need for classical interpolation and resampling. The new real-time linearization scheme has the additional capability of increasing the imaging acquisition speed by 10X while providing robustness to noise, properties that are demonstrated through mathematical analysis and simulation results throughout the paper.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.