Abstract

Combined the improved fast independent component analysis (FastICA) algorithm with the singular value decomposition algorithm, a single-channel fetal electrocardiogram (fECG) extraction method is proposed. First, the improved FastICA algorithm is used to estimate the maternal ECG component from a single-channel abdominal signal of pregnant women using an overrelaxation factor. Then, a preliminary estimate of the fECG signal is obtained by subtracting from the single-channel abdominal signal. Subsequently, the singular value decomposition algorithm is used to denoise the preliminarily estimated fECG signal to obtain a high signal-to-noise ratio. In addition, in the singular value decomposition algorithm for fetal arrhythmia, an improved method for constructing the ECG signal reconstruction matrix is proposed. Finally, the fECG extraction experiments on synthetic abdominal signals and actual abdominal signals (data from 49 abdominal channels sourced from DAISY database and the non-invasive fECG database in PhysioNet) are carried out. The experimental results show that the method in this paper can effectively improve the signal-to-noise ratio and the accuracy of fECG signal extraction, and is suitable for maternal or fetal arrhythmias. Compared with the FastICA algorithm, the signal-to-noise ratio of the fECG signal extracted by the method in this paper is improved by about 5 dB, and the accuracy of fECG extraction in the PhysioNet database can reach 96.54%.