Abstract

In a recent paper [Med. Phys. 19, 1441 (1992)] it was shown that the field size dependence of on-axis relative output from a linear accelerator can be well described by a model comprising two components. One component represents radiation emanating from a point source and the other represents radiation from a field size dependent distributed source. In this communication the application of the model is extended to off-axis situations and demonstrates that a good description of the dependence of off-axis output on collimator configuration can be obtained. Of particular clinical relevance in this regard is the beam profile of half blocked fields which are adequately reproduced. The model has also been applied to the study of profiles generated by simulated dynamic wedging.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.