Abstract

In the development of a high-precision vertical Fizeau interferometer with a 300-mm aperture, the deformation of the transmission flat because of clamping and gravity must be considered. In this paper, we proposed a ring-point support scheme for the deformation calibration of a large-diameter transmission flat. The calibration theory of the ring-point support system with elastic deformation was derived. The changes in the surface and stress field of the transmission flat were analyzed quantitatively by finite element method modeling, leading to the optimization of the support structure. To validate the proposed calibration approach, we performed an absolute test of the transmission flat using a liquid reference. The test result was compared to a measurement of the Zygo interferometer demonstrating the effectiveness of the proposed ring-point support design. Finally, with the iterative deformation calibration, the accuracy of the transmission flat reached λ/25 (Peak Valley, PV) for a 300-mm aperture.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.