Abstract

BackgroundMinimally invasive glaucoma surgeries (MIGS) have attracted significant attention, as they have been reported to lower intra-ocular pressure (IOP) and have an excellent safety profile. The iStent is an example of a minimally invasive glaucoma device that has received particular attention due to its early and wide spread utilization. There is a growing body of evidence supporting its use at the time of phacoemulsification to help lower IOP. However, it is still not clear how much of the IOP lowering effect can be attributed to the iStent, the crystalline lens extraction or both when inserted concurrently at the time of phacoemulsification. This has been an important issue in understanding its potential role in the glaucoma management paradigm.PurposeTo conduct a systematic review and meta-analysis comparing the IOP lowering effect of iStent insertion at the time of phacoemulsification versus phacoemulsification alone for patients with glaucoma and cataracts.MethodsA systematic review was conducted utilizing various databases. Studies examining the IOP lowering effect of iStent insertion in combination with phacoemulsification, as well as studies examining the IOP lowering effect of phacoemulsification alone were included. Thirty-seven studies, reporting on 2495 patients, met the inclusion criteria. The percentage reduction in IOP (IOPR%) and mean reduction in topical glaucoma medications after surgery were determined. The standardized mean difference (SMD) was computed as a measure of the treatment effect for continuous outcomes taking into account heterogeneity. Fixed-effect and random-effect models were applied.ResultsA 4% IOP reduction (IOPR%) from baseline occurred following phacoemulsification as a solo procedure compared to 9% following an iStent implant with phacoemulsification, and 27% following 2 iStents implants with phacoemulsification. Compared with cataract extraction alone, iStent with phacoemulsification resulted in significant reduction in the post-operative IOP (IOPR%) (SMD = -0.46, 95% CI: [-0.87, -0.06]). A weighted mean reduction in the number of glaucoma medications per patient was 1.01 following phacoemulsification alone compared to 1.33 after one iStent implant with phacoemulsification, and 1.1 after 2 iStent implants with phacoemulsification. Compared to cataract extraction alone, iStent with cataract extraction showed a significant decrease in the number of glaucoma medications (SMD = -0.65, 95% CI: [-1.18, -0.12]). Funnel plots suggested the absence of publication bias.ConclusionBoth iStent implantation with concurrent phacoemulsification and phacoemulsification alone result in a significant decrease in IOP and topical glaucoma medications. In terms of both reductions, iStent with phacoemulsification significantly outperforms phacoemulsification alone.

Highlights

  • Open angle glaucoma (OAG) is the leading cause of irreversible vision loss[1,2] affecting 60.5 million people globally

  • A 4% intraocular pressure (IOP) reduction (IOPR%) from baseline occurred following phacoemulsification as a solo procedure compared to 9% following an iStent implant with phacoemulsification, and 27% following 2 iStents implants with phacoemulsification

  • Compared with cataract extraction alone, iStent with phacoemulsification resulted in significant reduction in the post-operative IOP (IOPR%) (SMD = -0.46, 95% CI: [-0.87, -0.06])

Read more

Summary

Introduction

Open angle glaucoma (OAG) is the leading cause of irreversible vision loss[1,2] affecting 60.5 million people globally. The prevalence of OAG is projected to rise to nearly 80 million people by 2020.[3,4] OAG is an enormous public health burden, with the direct annual cost of treatment estimated at 2.86 billion (USD) in the U.S [5] and 300 million (CAD) in Canada [6] including productivity loss These costs are cumulative and can have a huge financial impact for patients, care givers, and for society at large.[7] the social consequences of severe OAG includes difficulties in performing daily activities, increased risk of falls[8], injuries, depression [9], hip fractures [10], and increased risk of death [1]. It is still not clear how much of the IOP lowering effect can be attributed to the iStent, the crystalline lens extraction or both when inserted concurrently at the time of phacoemulsification This has been an important issue in understanding its potential role in the glaucoma management paradigm.

Methods
Results
Conclusion