Abstract

To investigate the impact of with tympanostomy tubes (TT) on infrared tympanic membrane thermometer (ITMT) results and to provide a systematic review of ITMT results in non-naïve tympanic membranes. Original prospective blinded case series and systematic literature review. A single tertiary university-affiliated medical center. ITMT measurements of patients with unilateral TT and contralateral naïve control ear were randomly conducted by a single investigator blinded to the TT side before and after cerumen was removed from the external auditory canals. A systematic literature search of "MEDLINE" via "PubMed," "Embase," and "Google Scholar" on comparable published cases was performed. The mean paired differences (95% confidence interval [CI]) between ventilated and non-ventilated ears before and after cerumen removal were 0.08 ºC/0.14 ºF (-0.04 to 0.19 ºC/-0.07º-0.34º) and 0.62 ºC/1.12 ºF (0.04-0.25 ºC/0.07-0.45 ºF), respectively (P < 0.001 and P = 0.01, respectively). These findings support the validity and accuracy of ITMT in the setting of ventilated ears.