Abstract
3-Methylhistidine (3-MeHis) is increasingly used as an indicator of muscle protein breakdown. The development of asensitive, simple, and non-invasive method for 3-MeHis assay is important in clinical practice. Herein, a sensitive, simple, and non-invasive electrogenerated chemiluminescence (ECL) method was proposed for the quantitation of 3-MeHis in urine by using an iridium(III) solvent complex ([Ir(dfppy)2(DMSO)Cl], dfppy = 2-(2,4-difluorophenyl)pyridine, Ir-DMSO) as a signal reagent. The photoluminescence (PL) and ECL responses of Ir-DMSO to 3-MeHis were studied. The ECL intensity of Ir-DMSO was enhanced in the presence of 3-MeHis because of the coordination recognition between Ir-DMSO and the imidazole group of 3-MeHis. Based on the enhancement of ECL intensity, 3-MeHis can be sensitively detected in the range of 5 to 25μM. The detection limit was 0.4μM. This is the first report of an ECL method for the quantitation of 3-MeHis. Further, to investigate the feasibility of the Ir-DMSO-based ECL method in practical applications, the developed ECL method was applied for 3-MeHis assay in urine samples of 28 healthy volunteers and 2 patients. The urine samples from patients hospitalized with obesity and kidney disease and healthy individuals were distinguished by the ECL responses of Ir-DMSO. The proposed ECL method based on the coordination recognition between iridium(III) solvent complex and the imidazole group of 3-MeHis allows inexpensive, fast, non-invasive, and sensitive detection of 3-MeHis in urine, which is promising for assessing large volumes of patients for routine analysis in clinical practices.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have