Abstract

An LC–MS method has been developed combining ion-pair chromatography with an electrospray interface linking microbore and capillary HPLC to mass spectrometry. Separation of cyclic nucleotides on C 18 reversed-phase columns, using tetrabutylammonium bromide as an ion pairing agent was evaluated with different mobile phase compositions. It was found that low ion-pairing agent concentration (50–500 μ M) used in combination with low flow-rates (5–10 μl min −1) allowed the system to operate for up to several days without observing a reduced signal caused by source pollution. The loss of sensitivity expected in ion-pair chromatography could be remedied by using a 2-propanol coaxial sheath flow. Optimal conditions for negative ion electrospray resulted in a linear detection response in the femtomole to picomole range. Using biological samples this method was evaluated and compared with a classical ion-suppression RP-HPLC method using UV detection.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.