Abstract
This research investigates the utilization of an ionic liquid combination of solidified floating organic drop micro-extraction (IL-SFODME) to augment the concentration of trace amounts of lead, working as a preliminary stage before electrothermal atomic absorption spectrometry (ETAAS) analysis without the use of chelating agents. Key parameters impacting the microextraction efficiency-including pH, the volume of the ionic liquid (1-Hexyl-3-methylimidazolium hexafluorophosphate, HMIMPF6), temperature, extraction time, and stirring speed-were methodically examined to determine optimal conditions. Under detected optimized conditions, an enhancement factor of 71.2 was obtained for a 15 mL sample solution. The calibration curve exhibited linearity within the concentration range of 0.2-2.5 µg/L, with a detection limit (3σ) of 0.054 µg/L and a quantification limit (10σ) of 0.18 µg/L. For seven replicate measurements of 0.5 µg/L lead, the relative standard deviation (RSD) was ±2.30%. This method was effectively implemented to extract and quantify lead in both reference water and different real water samples, showcasing significantly efficient extraction performance.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have