Abstract
An oxalic acid-α-hydroxyisobutyric acid eluent has been used for the separation and determination of rare-earth elements by high-performance ion-exchange chromatography. Fifteen rare-earth elements were separated within less than 25 min on a 150×4.6 mm i.d. column packed with 5-μm sulfonic acid-bonded silica particles by elution with a combined gradient of 0.60–9.0 mM oxalic acid and 19.0–5.0 mM α-hydroxyisobutyric acid at pH 4.6. Detection and quantitation of the separated rare-earth elements was accomplished by visible-absorbance measurements at 600 nm after postcolumn reaction with arsenazo I. The gradient of the two complexing agents was optimized to enable the separation of yttrium(III) without interference from other elements, especially dysprosium(III) and terbium(III). Mass detection limits of the elements were in the range of 2–4 ng. Finally, the chromatographic system was applied to the quantitative analysis of rare-earth elements in monazite and xenotime.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.