Abstract

An ion-sensitive field effect transistor, which is a small potentiometric ion-sensing probe, was applied as a detector in the ion chromatography of alkali-metal cations. The cation-sensing transistor was prepared by coating the gate part of the probe to form a poly(vinyl chloride) matrix membrane containing tetranactin, which enables detection of alkali-metal ions such as Na+, K+, Rb+, and Cs+. To be able to analyse amounts less than 1 microliter and make full use of the small ion-sensing probe, we constructed a miniature cation-exchange separation column (support 10 microns polystyrene) with a PTFE tube (20 mm x 1.5 mm I.D. x 2.1 mm O.D). As an application of the system, Na+ and K+ concentrations in serum were determined. The analytical results for these two cations were good agreement with those obtained by flame photometry.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.