Abstract

A new technique for manipulating the kinetic energy distribution of electrospray ions that arrive at a Fourier transform ion cyclotron resonance trapped-ion cell is presented. Narrow kinetic energy distributions can complicate the selection of appropriate trapping conditions for electrospray ions and introduce charge discrimination in resulting mass spectra. Modulation of the applied skimmer potential controllably broadens the kinetic energy distribution, which improves the reproducibility of acquired spectra and eliminates charge discrimination. Mass spectra of horse heart cytochrome c are presented to demonstrate the utility of the technique. For example, applied static skimmer potentials of 12 and 9 V yield charge state distributions ranging from [M+19H](+19) to [M+12H](+12) and [M+15H](+15) to [M+7H](+7), respectively. A 12 ± 2 V, 100-Hz modulation of the skimmer potential yields an electrospray spectrum with charge states that range from [M+19H](+19) to [M+7H](+7), which is more representative of the source distribution.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.