Abstract
This paper presents an implantable package aimed at hosting a bidirectional neural interface for neural prosthetic applications. The package has been conceived to minimize the invasivity for the patient, for this reason a cylindrical container with an outer diameter of 7 mm and a length of 21 mm has been designed. The package, realized in alumina (Al2O3), presents 32 hermetic feedthroughs located at the top and bottom base of the cylinder. The hermetic housing has been assembled using a low-temperature soldering method based on a previous platinum/gold (Pt/Au) metallization of the ceramic parts. The package's hermeticity has been successfully proved by means of in-vitro tests, exhibiting an increase in the inner relative humidity of 20 %RH over 75 days of observation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.