Abstract

Thallium hydride was generated from aqueous solutions by merging sample and sodium tetrahydroborate reductant in a batch system. In situ preconcentration of volatile thallium hydride in a preheated graphite furnace coated with palladium, which was used as both the collection medium and atomizer, greatly improved the sensitivity for the determination of thallium by hydride generation atomic absorption spectrometry. The presence of tellurium can increase the generation efficiency of thallium hydride. The operating conditions were optimized. The calibration graph is linear up to 100 ng and the characteristic mass for thallium was 0.92 ng which is seventeen times lower than that obtained with the heated quartz tube atomizer.