Abstract

AbstractThe detection limits of a number of steroids have been determined by means of loop‐injection thermospray (TS) mass spectrometry. The measured limits were found to be in the range 1 pg‐1O ng/μL. Combined positive‐ion TS/tandem mass spectrometry yielded clear qualitative differences between mass selected protonatcd testosterone and estriol ions, both of which have the same nominal molecular weight. These differences are necessary for the correct identification of either species which may be present within a complex mixture. The heights of the peaks due to the two competing reaction channels, [M + H]+ and [M + NH4]+, were monitored as a function of the repeller voltage in the range 100‐350 V, and the implications of the acquired data are discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.