Abstract

Intraoperative neuromonitoring (IONM) of the recurrent laryngeal nerve (RLN) has recently been more frequently applied in thyroid surgery. However, concerns have been raised regarding the safety and optimal intensity of electrical nerve stimulation. Eight piglets were enrolled, and electrically evoked electromyography (EMG) was recorded from the vocalis muscles via endotracheal surface electrodes. The baseline EMG was measured and continuous pulsatile stimulations were performed on the vagus nerve and RLN for 10 minutes. Changes of EMG waveform and cardiopulmonary status were analyzed. A dose-response curve existed with increasing EMG amplitude as stimulating current was increased, with maximum amplitude elicited on vagal and RLN stimulation at <1 mA. No obvious EMG changes and untoward cardiopulmonary effects were observed after the stimulation. Electrical stimulation is safe during IONM in this porcine model. Minimal current that required generating the maximal evoked EMG, approximately 1 mA in this study, can be selected to minimize the risk of nerve damage and cardiopulmonary effects.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.