Abstract
In remote thyroid surgery, a reliable intraoperative neuromonitoring (IONM) procedure is an important tool for reducing injury to recurrent laryngeal nerve (RLN). This study proposes an alternative or adjunct technique for performing full percutaneous (PC) IONM and confirms its feasibility in animal experiments. This prospective porcine model study enrolled four piglets with eight nerve sides. Evoked electromyography (EMG) was stimulated from PC ball-tip probe, and recorded from EMG endotracheal tube (ETT) and from PC paired long-needle electrodes on the perichondrium of the lateral aspect of thyroid cartilage. In all RLNs and vagus nerves, typical laryngeal EMG waveforms were successfully evoked by PC probe stimulation and recorded by both ETT and PC needle electrodes. This study confirms the feasibility of the full PC IONM techniques in porcine model. However, further clinical studies are needed to compare the practicality of different remote-access approaches for thyroid surgery.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.