Abstract

Practical aspects of application of angular ultrasonic transducers for measurement parameters of multi-layered structures are analyzed. The algorithms for determination of thickness of every separate layer in the multi-layered structure, by the use of angular ultrasonic transducers as well as by the use of longitudinal or shear waves, are proposed. It is shown that the ratio of thickness of the first layer to the transversal dimensions of angular ultrasonic transducers is very important when the angle of incidence is selected and the parameters of multi-layered structures are determined. The algorithms for determination of minimal thickness of the first layer, which may be determined by the use of angular ultrasonic transducers with a given transversal dimensions, are developed. The modeling of dependence of the ratio of minimal thickness of the first layer to the transversal dimensions of angular transducers in the plane of incidence was performed, when the angle of incidence and propagation velocity of ultrasound signals in the first layer of structure are changed. The results of modeling are presented and analyzed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.