Abstract

Elekta AQUA v2.02 software (Gantry Runout isocenter test) was investigated as a tool for verification of kilovoltage to megavoltage gantry radiation isocenter coincidence. AQUA reported megavoltage (6 MV) isocenter was independent of field size over the range 5cm × 5cm to 20cm × 20cm. For the 10cm × 10cm field size, standard deviation in AQUA reported 3D megavoltage (6MV) isocenter over ten consecutive deliveries was less than 0.04mm for any axis. Compared to the full AQUA test delivery (Gantry Runout), the shorter AQUA test version (Gantry Runout short) gave a root mean square MV isocenter (± 1 SD) difference of 0.18mm ± 0.08. Across 7 machines, root mean square differences between AQUA and PIPS PRO reported MV isocenter (for 6 MV and 6 MV FFF beams) was 0.1mm ± 0.1mm, with most of the difference observed in the gun-target (Y-axis). AQUA 6 MV isocentre position was offset to gantry relative to Elekta XVI customer acceptance test (CAT) workflow by 0.25mm to 0.52mm. For 6 MV FFF beams, AQUA reported an MV isocenter of between 0.27 and 0.39mm offset to target relative to Elekta CAT.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.