Abstract

Wear and corrosion in total hip replacement negatively impact implant service-life and patient well-being. The aim of this study was to generate a statistical response surface of material loss using an apparatus, capable of testing the effect of wear and corrosion products in situ on cells, such as macrophages. The test chamber of a ball-on-flat tribometer operating inside a CO2 incubator was integrated with an electrochemical setup and adapted for cell culture work. A 20-test series, following a 2-level 3-factor design of experiments, was performed with a ceramic head in reciprocating rotational motion against a CoCrMo-alloy disc, under constant load. The lubricant was cell culture medium (RPMI-1640 + 10 vol% bovine serum). Response surfaces were generated, which statistically showed the influence of motion amplitude, load, and potential on the total mass loss and wear scar volume of the metallic discs. Potential had the highest impact on the total mass loss, while motion amplitude and load significantly influenced the wear scar volume. The concentrations of the alloy elements found in the lubricants reflected the bulk-alloy stoichiometry. The total concentration of Co released into the lubricant (2.3–63 ppm by total mass loss, 1.5 to 62 ppm by ICP-MS) corresponded well with the known range to trigger cell response. Tribocorrosion tests in the presence of cells and tissues, such as macrophages, lymphocytes and/or synovium, will be carried out in the future.