Abstract

To investigate the mechanical properties and wear of nylon inserts and abutments in Locator-retained, implant-supported overdentures (L-IODs). Clinical wear of inserts and abutments was qualitatively rated in a group of 16 patients with L-IODs. The inserts were also subjected to microscopic analysis, differential scanning calorimetry (DSC), and thermal gravimetric analysis (TGA). Wear was identified in almost all inserts and abutments. These results were corroborated by DSC and TGA analyses, which showed significant mechanical deterioration of the inserts. Nylon inserts and Locator abutments show relevant signs of deterioration in clinical use, indicating that regular maintenance is an issue that should be addressed with the patients prior to treatment.