Abstract

AbstractAim:To investigate the central electrode artefact effect of different ion chambers in the verification phantom using the dose calculation algorithms Analytical Anisotropic Algorithm (AAA) and Acuros XB.Materials and methods:The dosimetric study was conducted using an in-house fabricated polymethyl methacrylate head phantom. The treatment planning system (TPS)-calculated doses in the phantom with detectors were compared against the dummy detector fillets using AAA and Acuros XB algorithm. The planned and measured doses were compared for the study.Results:The mean percentage variation in volumetric-modulated arc therapy plans using Acuros XB and the measurement in the head phantom are statistically significant (p-value = 0.001) for FC65 and CC13 chambers. In small volume chambers (A14SL and CC01), the measured and TPS-calculated dose shows a good agreement.Findings:The study confirmed the CT set of the phantom with detectors (FC65 and CC13) give more artefacts/heterogeneity caused a significant variation in dose calculation using Acuros XB. Therefore, the study suggests a method of using phantom CT set with the dummy detector for mean dose calculation for the Acuros XB algorithm.