IntroductionStereotactic body radiation therapy (SBRT) was found effective in treating laryngeal cancer with only five treatment fractions by a recent clinical trial (NCT01984502, ClinicalTrials.gov). Nevertheless, this trial used the Cyberknife system, which is not widely accessible enough to benefit all patients affected by laryngeal cancer. Our study investigates the feasibility of larynx SBRT treatment planning on a conventional gantry-based LINAC and compares its plan quality with that from the Cyberknife.Materials & methodsTen larynx SBRT cases were originally treated by Cyberknife using fixed cones in our institution, with plans created and optimized using the Monte-Carlo algorithm in the MultiPlan treatment planning system. These cases were retrospectively planned in the Eclipse planning system for a LINAC with the same prescription dose. We used volumetric modulated arc therapy (VMAT) for larynx SBRT planning in Eclipse and incorporated non-coplanar arcs to approach the Cyberknife’s large solid angle delivery space. We used both anisotropic analytical algorithm (AAA) and Acuros XB (AXB) algorithm for dose calculation and compared their accuracy by measurements on an in-house larynx phantom. We compared the LINAC VMAT plans (VMAT-AAA and VMAT-AXB) with the original Cyberknife plans using dosimetric endpoints such as the conformity index, gradient indices (R50, R20), OAR maximum/mean doses, and the monitor units.ResultsPhantom measurement showed that both the AAA and the AXB algorithms provided adequate dose calculation accuracy (94.7% gamma pass rate on 2%/2 mm criteria for AAA vs. 97.3% for AXB), though AXB provided better accuracy in the air cavity. The LINAC-based VMAT plans achieved similar dosimetric endpoints as the Cyberknife planning, and all plans met the larynx SBRT dosimetric constraints. Cyberknife plans achieved an average conformity index of 1.13, compared to 1.20 of VMAT-AXB and 1.19 of VMAT-AAA. The VMAT plans spared the thyroid gland better with average Dmean of 2.4 Gy (VMAT-AXB) and 2.7 Gy (VMAT-AAA), as compared to 4.3 Gy for Cyberknife plans. The VMAT-AAA plans had a slightly lower contralateral arytenoid Dmax (average: 15.2 Gy) than Cyberknife plans (average: 17.9 Gy) with statistical significance, while the contralateral arytenoid Dmax was similar between VMAT-AXB and Cyberknife plans with no statistically significant difference. Cyberknife plans offered slightly better R50 (average: 5.0) than VMAT-AXB (5.9) and VMAT-AAA (5.7) plans. The VMAT plans substantially reduced the plan MUs to less than 1/3 of the Cyberknife plans, and the differences were statistically significant. The other metrics were similar between VMAT and Cyberknife plans with no statistically significant differences.ConclusionsGantry-based LINACs can achieve similar plan quality to Cyberknife systems. Treatment outcome with both methods remains to be investigated.

Highlights

  • Stereotactic body radiation therapy (SBRT) was found effective in treating laryngeal cancer with only five treatment fractions by a recent clinical trial (NCT01984502, ClinicalTrials.gov)

  • Under a stricter gamma criterion (2%/2 mm), our analysis still yielded over 97.3% pass rate for Acuros XB (AXB) and 94.7% for analytical algorithm (AAA)

  • The results showed that VMATAXB achieved plan quality similar to volumetric modulated arc therapy (VMAT)-AAA after optimization, even though the air cavity posed additional challenges for planning target volume (PTV) coverage and OAR sparing

Read more

Summary

Introduction

Stereotactic body radiation therapy (SBRT) was found effective in treating laryngeal cancer with only five treatment fractions by a recent clinical trial (NCT01984502, ClinicalTrials.gov). This trial used the Cyberknife system, which is not widely accessible enough to benefit all patients affected by laryngeal cancer. Some previous studies have investigated the feasibility of reducing the total number of fractions of larynx radiotherapy to 25–28, and they have achieved equivalent or even better local control rates without increasing toxicity [5, 6] These promising results suggest further hypofractionation may achieve better local control and help to minimize the treatment length and cost

Methods
Results
Discussion
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.