Abstract

Homodyne and heterodyne laser interferometers can be used for dynamic force calibration and for acceleration measurement with sinusoidal and with shock excitation. The characterisics of both types of interferometers in acceleration measurement are compared in this paper. The data acquisition system comprises a four channels ADC card and a digital multimeter to sample the interferometer and force transducer or acccelerometer signals synchronously. The nonlinearity effects of homodyne and heterodyne interferometers in acceleration measurement are analyzed. A transducer is calibrated using both interferometers and good accordance is demonstrated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.