Abstract
In this paper, influence coefficient of flux (αi,f) on X-ray intensity was proposed, by which the X-ray intensity of fusion beads with different dilution ratio can be corrected to the standard value. Hence, the unknown sample of random dilution ratio can be determined by the standard working curve after corrected withαi,f. It was found that αi,f obtained by the proposed experimental method can be applied in correction of fluorescence intensity in the routine analytical work with satisfactory results. There is no necessary of exact weighing for the sample and the flux with the correction of αi,f. Furthermore, the procedure of routine analytical work can be simplified greatly without getting the L.O.I ahead of weighing, by which the working efficiency can be improved as well.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.