Abstract

We have successfully proposed a simulation of a neutron beam-shaping assembly using MCNPX Code. This simulation study deals with designing a compact, optimized, and geometrically simple beam shaping assembly for a neutron source based on a proton cyclotron for BNCT purpose. Shifting method was applied in order to lower the fast neutron energy to the epithermal energy range by choosing appropriate materials. Based on a set of MCNPX simulations, it has been found that the best materials for beam shaping assembly are 3 cm Ni layered with 7 cm Pb as the reflector and 13 cm AlF3 the moderator. Our proposed beam shaping assembly configuration satisfies 2 of 5 of the IAEA criteria, namely the epithermal neutron flux 1.25 × 109 n.cm-2 s-1 and the gamma dose over the epithermal neutron flux is 0.18×10 -13 Gy.cm 2 n -1. However, the ratio of the fast neutron dose rate over neutron epithermal flux is still too high. We recommended that the shifting method must be accompanied by the filter method to reduce the fast neutron flux.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.