Abstract
The intravoxel incoherent motion (IVIM) model extracts both functional and structural information of a tissue using motion-sensitizing gradients. The Objective of the present work is to investigate the impact of signal to noise ratio (SNR) and physiologic conditions on the validity of IVIM parameters. This study is a simulation study, modeling IVIM at a voxel, and also done 10,000 times for every single simulation. Complex noises with various standard deviations were added to signal in-silico to investigate SNR effects on output validity. Besides, some blood perfusion situations for different tissues were considered based on their physiological range to explore the impacts of blood fraction at each voxel on the validity of the IVIM outputs. Coefficient variation (CV) and bias of the estimations were computed to assess the validity of the IVIM parameters. This study has shown that the validity of IVIM output parameters highly depends on measurement SNR and physiologic characteristics of the studied organ. IVIM imaging could be useful if imaging parameters are correctly selected for each specific organ, considering hardware limitations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.