Abstract

BackgroundTo determine the reliability and usefulness of intraoperative monitoring of the abducens nerve during extended endonasal endoscopic skull base tumor resection. MethodsWe performed abducens nerve intraoperative monitoring in 8 patients with giant clival lesions recording with needle electrodes sutured directly into the lateral rectus muscles of the eye to evaluate spontaneous electromyographic activity and triggered responses following stimulation of the abducens nerves. ResultsA total of 16 abducens nerves were successfully recorded during endoscopic endonasal skull base surgeries. Neurotonic discharges were seen in two patients (12% [2/16] abducens nerves). Compound muscle action potentials of the abducens nerves were evoked with 0.1–4mA and maintained without changes during the neurosurgical procedures. No patient had new neurological deficits or ophthalmological complications post-surgery. ConclusionsIntraoperative monitoring of the abducens nerve during the extended endonasal endoscopic approach to skull base tumors appears to be a safe method with the potential to prevent neural injury through the evaluation of neurotonic discharges and triggered responses.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.