Abstract

The purpose of this study was to measure intraocular pressure (IOP) elevation while applying standard gonioscopy, selective laser trabeculoplasty (SLT), and laser iridotomy procedural lenses. Twelve cadaver eyes were mounted to a custom apparatus and cannulated with a pressure transducer which measured IOP. The apparatus was mounted to a load cell which measured the force on the eye. Six ophthalmologists performed simulated gonioscopy (Sussman 4 mirror lens), SLT (Latina lens), and laser iridotomy (Abraham lens) while a computer recorded IOP (mm Hg) and force (grams). The main outcome measures were IOP and force applied to the eye globe during ophthalmic diagnostics and procedures. The average IOP's during gonioscopy, SLT, and laser iridotomy were 43.2 ± 16.9 mm Hg, 39.8 ± 9.9 mm Hg, and 42.7 ± 12.6 mm Hg, respectively. The mean force on the eye for the Sussman, Latina, and Abraham lens was 40.3 ± 26.4 grams, 66.7 ± 29.8 grams, and 65.5 ± 35.9 grams, respectively. The average force applied to the eye by the Sussman lens was significantly lower than both the Latina lens (P = 0.0008) and the Abraham lens (P = 0.001). During gonioscopy indentation, IOP elevated on average to 80.5 ± 22.6 mm Hg. During simulated laser iridotomy tamponade, IOP elevated on average to 82.3 ± 27.2 mm Hg. In cadaver eyes, the use of standard ophthalmic procedural lenses elevated IOP by approximately 20 mm Hg above baseline.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call